Start the Pulse tube from Control panel. The system cools to 12 Kelvin after 10 hours.
Start to pump Mixture in and out through the Still line: this can be done with the "pulse_precool" script. Wait for 1-2 hours until P3 is few tens of mbar. Stop the script at the pumping stage (V10 is open)
Stop the heatswitches
Put the cryostat into the normal circulation mode (V10, scroll1, V8, V6, Com.,V5 and V1 on/open)
Start the condense_v2 script, flow should be around 1 mmol/s